Model Number N/A
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Patient Involvement (2645)
Event Date 12/08/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per getinge standard operating procedure since the iabp was manufactured more than a year before the date of event.A getinge field service engineer (fse) was dispatched to evaluate this unit and was unable to reproduce the reported failure.The unit functioned with test balloon.The fse found that the, helium tank was closed.The unit had 300 psi at start of checks.The customer had unit set to external for ecg and blood pressure readings.The unit passed all calibration, functional and safety tests per factory specifications and was returned to customer and cleared for clinical use.
 
Event Description
It was reported that prior to use on a patient the cs300 intra-aortic balloon pump (iabp) generated several augmentation alarms and autofill failure #57.There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8206078
MDR Text Key131945275
Report Number2249723-2018-02239
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-