| Catalog Number |
01985 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The review of the batch manufacturing records was conducted, and the batch met all finished goods release criteria.To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.Attempts are being made to obtain the additional information.To date no response has been provided.If further details are received at the later date a supplemental medwatch will be sent.It was reported the product broke.Please explain what part of the device broke and how did it brake? did the broken piece fall into the patient? was the broken piece retrieved? was there any patient consequence or any medical/surgical intervention and/or treatment rendered? will any of the product be returned for analysis?.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a hysteroscopy procedure on (b)(6) 2018 and the electrosurgical device was used.It was also reported that the device broke.There was no patient consequence reported.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Date sent to the fda: 03/04/2019.Additional summary: the device has not been returned in its original packaging.The active tip is snapped and deformed.The device shaft has bend in the middle of the shaft.The device as presented appears to have seen a level of misuse to the shaft and the distal active tip is snapped in the centre position.It is not known how the damage to the shaft has occurred, it may have been due to shipping without protection.The condition of the fracture face of the active tip suggests a brittle failure with no obvious signs of mechanical damage.Instances of similar failures have been seen historically, with the failure location, mode and method indicating failure due to surface (hydrogen) embrittlement.Ifu advise the user not to operate the electrode in the beak of the sheath.Cause code: user / use error | wrong handling; user / use error | excessive force.
|
| |
|
Search Alerts/Recalls
|