| Model Number |
238131 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product involved in the incident was returned to arkray usa and forwarded to the manufacturer for evaluation.We have not yet received evaluation results.Arkray will file a follow-up report when device evaluation is complete.
|
| |
|
Event or Problem Description
|
|
Customer is using the techlite pen needles and have been using the brand for one year.He stated when he puts the pen needle on to his lantus, it doesn't pierce the device right away and causes the needle to bend.He is using part (b)(4).Verified that the customer is not re-using the pen needle.It is bending on the first use.He is stating that it's very difficult to get the needle on to the lantus.They must apply pressure.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhrs were reviewed and no records confirming the defect were observed.Archival and returned samples meet requirements of specification.Additionally, 10 randomly selected pen needles from archival and returned samples were assembled on pen injector.With every attempt, droplets on the cannula were observed and injection of applied dosage can be seen which confirms the correctness of the pen needle assembly on the pen and the patency of the needles.Defect was not confirmed.Root cause analysis was not conducted.No corrective action.Complaint closed.
|
| |
|