• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EMP STEM 9 SO 140MM; PRSTHSIS,HIP,SMI-CNSTRAIND,UNCMNTD,METAL/POLYMR,NON-POROUS, CALICUM-PHOSPHATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EMP STEM 9 SO 140MM; PRSTHSIS,HIP,SMI-CNSTRAIND,UNCMNTD,METAL/POLYMR,NON-POROUS, CALICUM-PHOSPHATE Back to Search Results
Catalog Number 71290902
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Injury (2348)
Date of Event 12/04/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to periprosthetic fracture/implant failure.Due to where the fracture of the prosthesis was, couldn't use emperion removal instruments and removal of stem was difficult as the distal stem was well fixed.Then cut base of sleeve however was unable to remove it.Then tried the trephines over the stem and still unable to free the stem.So an extended trochanteric osteotomy was performed, strut graft and the longest accord trochanteric grip plate was used over the restoration hip system with 9 cables in total.The liner was not replaced.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.The clinical/medical team concluded, the x rays provided confirm the broken stem and possible stress shielding around the greater trochanter which would indicate lack of support of the neck of the stem.In addition, the product evaluation confirmed the stem breakage was due to fatigue.However, without the revision report, or the post implantation radiographs we cannot rule out misalignment as a possible contributing factor.The impact to the patient beyond the revision cannot be determined.Should additional information become available this issue can be re elevated.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re opened.Mimb review.Assess severity of complaint case to determine if additional actions or inputs are required for inclusion in the medical assessment.Determine if a medical assessment will be performed based on a review of the complaint details and further input from the medical director/designee.Reviewed during mimb.A medical investigation will be performed.Proceed based on information provided/available for the investigation; if no relevant clinical information is provided, recommend closure.Approved by (b)(6), medical director.A review of relevant clinical/medical information in the reported issue, inclusive of technique and patient information, to include, but not limited to: patient information.Surgical procedure/post-operative care review.Device labeling (including technique guides, ifus, etc.).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMP STEM 9 SO 140MM
Common Device Name
PRSTHSIS,HIP,SMI-CNSTRAIND,UNCMNTD,METAL/POLYMR,NON-POROUS, CALICUM-PHOSPHATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8206177
Report Number1020279-2018-02952
Device Sequence Number19323036
Product Code MEH
UDI-Device Identifier03596010561275
UDI-Public03596010561275
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K052426
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71290902
Device Lot Number06GM13128
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer06/05/2019
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date06/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71343204/09BM02196; 71343204/09BM02196
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-