MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HI FLEX ISRT SZ 3-4 15; PRSTHSIS,KNEE,PATELLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HI FLEX ISRT SZ 3-4 15; PRSTHSIS,KNEE,PATELLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/METAL/POLYMER Back to Search Results
Catalog Number 71421511
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 12/05/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right tibial baseplate and the ps tibial insert were revised due to pain.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A review of the attached photos does show significant wear on these devices.The devices were manufactured in 2008.A clinical evaluation was conducted and this case reports that a revision surgery was performed due to pain.The initial implantation occurred 10 years prior to the revision.During the revision the right tibial baseplate and ps tibial insert was revised.Two undated x-rays were provided and the components appear well positioned with maybe some radiolucency around the femoral component but can¿t be confirmed if loosening without the revision report.Details regarding the patient¿s weight-bearing status, medical history, bone quality, fall/trauma history, and other additional clinical relevant information have not been provided.Based on the limited information provided and without the return of the device for evaluation the root cause of the reported pain cannot be determined.Should additional information become available this issue can be re-elevated.No further clinical assessment is warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT SZ 3-4 15
Common Device Name
PRSTHSIS,KNEE,PATELLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8206209
Report Number1020279-2018-02953
Device Sequence Number16619508
Product Code JWH
UDI-Device Identifier03596010492937
UDI-Public03596010492937
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/09/2018
Device Catalogue Number71421511
Device Lot Number08AT16690
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer12/05/2018
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date03/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-