Catalog Number ASKU
Device Problem Material Discolored (1170)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported that several whitish submillimeter specks appeared in the anterior chamber of a patients right eye during the use of irrigation-aspiration.The physician immediately removed the visible specks except one that was unattainable as it was placed behind the lens implant.The surgeon reports that it is unknown if the particle caused any damage to the eye.
 
Manufacturer Narrative
As the customer did not retain the finished goods lot number, device history record (dhr) and lot history could not be reviewed.The cassette was not returned, therefore, visual or functional testing could not be conducted in order to ascertain a potential failure mode of the consumable device.The root cause of the customer's complaint could not be established as a sample has not been received.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.After an investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned and a root cause evaluation could not be performed.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key8206227
MDR Text Key131744191
Report Number1644019-2018-00295
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received01/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BALANCED SALT SOLUTION; CENTURION VISION SYSTEM; ULTRAFLOW II I/A HANDPIECE; UNSPECIFIED VISCOELASTIC; BALANCED SALT SOLUTION; CENTURION VISION SYSTEM; ULTRAFLOW II I/A HANDPIECE; UNSPECIFIED VISCOELASTIC
Patient Age75 YR
-
-