• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD¿ INTERLINK® LEVER LOCK CANNULA; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD¿ INTERLINK® LEVER LOCK CANNULA; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 303370
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device expiration date: unknown.(b)(6).Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd¿ interlink® lever lock cannula leaked blood after injection.
 
Additional Manufacturer Narrative
Investigation summary: three photos and one loose lever lock sample were received and visually evaluated.The photos depict a loose lever lock body next to a bagged injection tubing apparently filled with blood.The lever lock sample was visually inspected.It was confirmed to be from mold/cavity x-86.No molding defects were found in any parts of the sample returned.Lot number is unknown; therefore dhr could not be performed.No visual defects were observed in the lever lock body.The lever lock sample was tested for leakage.No leakage was found at any point within the body of the sample.The reported defect was not identified in the samples received.
 
Event or Problem Description
It was reported that the bd¿ interlink® lever lock cannula leaked blood after injection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ INTERLINK® LEVER LOCK CANNULA
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key8206278
Report Number1213809-2018-00948
Device Sequence Number16619658
Product Code FPA
UDI-Device Identifier30382903033707
UDI-Public30382903033707
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K940697
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number303370
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2018
Initial Date Received by Manufacturer 12/13/2018
Supplement Date Received by Manufacturer12/13/2018
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date04/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-