• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VNMC4040C55TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 12/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant navion stent graft was implanted in a patient for the endovascular treatment of a thoracic aortic ulcer.It was reported that the patient had a good 1 week post operative scan, however presented with an ascending aortic dissection at the 1 month scan.The dissection was visible proximal to and behind the implanted device.The patient has not received treatment, as the physician opted for observation.As per the physician, the cause of the event may be device oversizing, but feels the issue relates to the imh and weak tissue in the patient's anatomy.No additional clinical sequelae were reported and the patient will be monitored.
 
Additional Manufacturer Narrative
Additional information: it was reported that the ulcer was located proximal to the lsa and the proximal end of the stent graft was implanted in zone 0 of the ascending aorta.It was reported that 10-20% stent graft oversizing was used.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8206317
Report Number9612164-2018-03790
Device Sequence Number2069306
Product Code MIH
UDI-Device Identifier00763000101084
UDI-Public00763000101084
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/07/2020
Device Model NumberVNMC4040C55TU
Device Catalogue NumberVNMC4040C55TU
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/18/2018
Supplement Date Received by Manufacturer01/17/2019
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age88 YR
-
-