Catalog Number 0066214
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess the instrument test well images in question on m00570, and determined that they appeared visually negative.No errors were noted in the test event log for the blood sample in question, and the camera report was acceptable.For the m00569 instrument, immucor technical support determined that the test wells in question appeared visually negative.No errors were noted in the test event log for the blood sample in question, and the camera report was acceptable.The immucor laboratory tested retention product on 12dec2018, which performed as expected.The internal immucor tracking record number is (b)(4).
 
Event or Problem Description
On (b)(6) 2018, a customer site reported an unexpectedly negative antibody identification when using capture-r ready-id, when used on a galileo echo instrument, when tested on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-ID
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key8206417
Report Number1034569-2018-00284
Device Sequence Number19362560
Product Code KSZ
UDI-Device Identifier10888234001188
UDI-Public10888234001188
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date01/08/2019
Device Catalogue Number0066214
Device Lot NumberID369
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2018
Initial Report FDA Received Date12/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-