• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Disconnection (1171)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/28/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 8709sc, serial#: (b)(4), implanted: (b)(6) 2012, product type: catheter.Other relevant device(s) are: product id: 8709sc, serial/lot #: (b)(4), ubd: 24-may-2014, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care professional (hcp) via a device manufacturer representative regarding a patient receiving morphine (40 mg/ml at 13.998 mg/day) and clonidine (1400 mcg/ml at 489.6 mcg/day) via an implantable infusion pump.The indication for use was noted to be non-malignant pain.It was reported that during a normal and expected pump replacement for battery depletion it was determined that the catheter was not connected to the pump.The surgery was on (b)(6) 2018.There were no patient symptoms reported and no issues noted prior to the surgery.The cause of the event was undetermined.The catheter was securely attached to the new pump.The issue was considered resolved.The patient's weight was provided.The patient's status at the time of the report was alive - no injury.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8206580
MDR Text Key131782539
Report Number3004209178-2018-28516
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00613994779229
UDI-Public00613994779229
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2013
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2018
Initial Date FDA Received12/31/2018
Date Device Manufactured03/02/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient Weight41
-
-