MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH SPINDLE FOR EVOLUTION; MISC ORTHO SURGICAL INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH SPINDLE FOR EVOLUTION; MISC ORTHO SURGICAL INSTR Back to Search Results
Catalog Number 03.825.002
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is synthes sales consultant.A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that on an unknown date, during a spinal procedure of an internal stem extremity of the two (2) trial implant holder synfix / syncage evolution spindle were unscrewed so it was hard to remove an unknown trial implant.There was no patient consequence reported.It was unknown how the procedure was completed and if there was any surgical delay.Patient outcome was unknown.Concomitant device: trial implant (part unknown, lot unknown, quantity 1).This report is for one (1) spindle for evolution.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Visual inspection: upon visual inspection, the knob was received separated from the spindle.It was observed that the weld around the knob was cracked around the entire circumference.The superior surface of both knobs show visible signs of impaction consistent normal use.Therefore, this complaint is confirmed.Dimensional inspection: since the complaint condition involved a cracked weld, a relevant dimensional check could not be performed.Document/specification review: review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Conclusion: previous investigations on the complaint condition determined that the weld was not strong enough to withstand the applied impaction load from normal use.Therefore, during the investigation no unidentified product design issues or discrepancies were observed that may have contributed to the complaint condition.While no definitive root cause could be determined it is possible that the complaint condition is the result of force applied on the device resulting in stresses beyond the failure limit during usage.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot manufacturing location: supplier - avalign, inspected, packaged and released by: monument release to warehouse date: 16-may-2017 part number: 03.825.002, spindle for evolution lot number: h272722 (non-sterile) this lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINDLE FOR EVOLUTION
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8206649
Report Number8030965-2018-59341
Device Sequence Number13970562
Product Code LXH
UDI-Device Identifier07611819542303
UDI-Public(01)07611819542303
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.825.002
Device Lot NumberH272722
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Initial Date Received by Manufacturer 12/06/2018
Supplement Date Received by Manufacturer01/11/2019
02/21/2019
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date01/30/2019
02/25/2019
Patient Sequence Number1
-
-