Catalog Number BMK6F95BER120 |
Device Problems
Device Damaged Prior to Use (2284); Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a medical procedure, the physician found that the benchmark 6f 071 delivery catheter (benchmark) was kinked at the hub, upon removal from the packaging loop.The damage to the benchmark was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using another benchmark.
|
|
Manufacturer Narrative
|
Results: the benchmark was kinked approximately 5.0 cm and 47.0 cm from the hub.Bends were present approximately 15.0 cm, 21.0 cm and 38.0 cm from the hub.The device was ovalized approximately 95.0 cm and 96.0 ¿ 100.0cm from the hub.Conclusions: evaluation of the returned benchmark confirmed the device was kinked.If the device is removed from the packaging tube at extreme angles, damages such as kinks are likely to occur.Further evaluation revealed ovalizations and bends along the device.These bends and ovalizations are likely incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|