Catalog Number 368861
Device Problems Short Fill (1575); Volume Accuracy Problem (1675)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event Description
It was reported that bd vacutainer® k2e (edta) 7.2mg tubes do not fill completely.
 
Event Description
It was reported that 50 bd vacutainer® k2e (edta) 7.2mg tubes do not fill completely.
 
Manufacturer Narrative
The following field has been updated due to corrected information: describe event or problem: it was reported that 50 bd vacutainer® k2e (edta) 7.2mg tubes do not fill completely.Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for underfill with the incident lot was observed.Additionally, retention samples were selected from bd inventory for evaluation/testing and upon completion, the issue relating to underfill was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for underfill with the incident lot was observed.Additionally, evaluation/testing of the retain samples was conducted and underfill was not observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Rationale: based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2E (EDTA) 7.2MG TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
MDR Report Key8207043
MDR Text Key131763351
Report Number9617032-2018-03245
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903688611
UDI-Public50382903688611
Combination Product (y/n)N
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2019
Device Catalogue Number368861
Device Lot Number8040575
Initial Date Manufacturer Received 12/11/2018
Initial Date FDA Received12/31/2018
Supplement Dates Manufacturer Received12/11/2018
Supplement Dates FDA Received02/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-