| Model Number |
ZCT150 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct150, 23.0 diopter intraocular lens (iol) when being inserted into the eye of the patient, a scratch was noticed on the iol.The lens was removed and replaced with the back-up lens.Patient was fine post-op.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product testing could not be performed because the product was not returned.The reported issue could not be verified.Manufacturing record review: the manufacturing process record was evaluated and there were no non conformances with respect to the manufacturing process.The product was manufactured and released according to specifications.A search revealed that no additional complaints were received from this production order.Labeling review: the labeling review was completed.The directions for use (dfu) adequately provide instructions, precautions, along with warnings for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|