Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server.In this instance, there was no evidence that the physician received the message warning that the report did not upload.It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved.Troubleshooting from unity support found that there was a reportid generated, however the document was not uploaded to i:\ drive.A new word document was created and renamed reportid.The exam was re-read and confirmed with the customer that the report was re-read without further issues.
|
|
Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2018 the customer reported an exam was completed, however, there is no exam report.An investigation showed the report did not upload successfully and had to be read again by the doctor.There was no reported adverse event to the patient.However, while images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm.(b)(4).
|