| Model Number |
MMT-723NAP |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
12/29/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose level of 400 mg/dl.The customer did not provide the symptoms regarding high blood glucose.The customer was stated that the insulin pump alarmed with motor error.The insulin pump will not be returned for analysis.
|
| |
|
Search Alerts/Recalls
|