C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX¿ SELF-ADHERING MALE EXTERNAL CATHETER; MEC Back to Search Results
Model Number 33302
Health Effect - Clinical Codes Pain (1994); Skin Irritation (2076)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." the device was not returned.
 
Event or Problem Description
It was reported that the adhesive of the male external catheter was too strong and left residue on the patient.The patient allegedly experienced pain and skin irritation.No medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX¿ SELF-ADHERING MALE EXTERNAL CATHETER
Common Device Name
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8208531
Report Number1018233-2019-00005
Device Sequence Number19552952
Product Code NNX
UDI-Device Identifier00801741070754
UDI-Public(01)00801741070754
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number33302
Device Catalogue Number33302
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2018
Initial Report FDA Received Date01/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-