| Catalog Number |
180704-1 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There has been one other similar event for the lot referenced which occurred during the same procedure for the same patient.Material analysis: damage observed on insert consistent with attempted implantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
During the final stage of the case the surgeon was attempting to place the poly.He was concerned about the poly not sitting down on the lateral side of the base plate.After several attempts to get the poly to seat, i had someone run and grab a second poly.We opened that and had the same issue.I noticed that the new poly shared a lot code with the original.When we could not get the second poly with the same lot code to seat, we opened a third implant that had a different lot code.This poly went in and sat on the tray like it should.
|
| |
|