ENCORE MEDICAL L.P. FOUNDATION PRIMARY KNEE/LATERAL PIVOT INSERT; INSERT, STD., PRIMARY, SIZE8 9MM CM Back to Search Results
Model Number 385-09-508
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 12/11/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient presenting multiple times to the doctor complaining of pain.Doctor considered instability.Operatively the surgeon found nothing of concern; replaced the poly since it had been disrupted.The doctor now doubts if the patient had pain.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to pain.The previous surgery and the surgery detailed in this investigation occurred 6.9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION PRIMARY KNEE/LATERAL PIVOT INSERT
Common Device Name
INSERT, STD., PRIMARY, SIZE8 9MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8210014
Report Number1644408-2018-01188
Device Sequence Number13959970
Product Code JWH
UDI-Device Identifier00888912117760
UDI-Public(01)00888912117760
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K000590
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/07/2022
Device Model Number385-09-508
Device Catalogue Number385-09-508
Device Lot Number485B1044
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2018
Supplement Date Received by Manufacturer01/29/2019
Initial Report FDA Received Date01/02/2019
Supplement Report FDA Received Date01/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-