• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Use of Device Problem (1670)
Patient Problems Hypoglycemia (1912); Ambulation Difficulties (2544); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
The pump has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2018, the reporter contacted animas alleging that the patient experienced a blood glucose between 40-50 mg/dl with sweating, shakes and unsteadiness when standing and/or walking associated with an alleged history/settings issue.Reportedly, the patient remained on the pump and did not receive any treatment above and beyond the usual routine of diabetes care and management.During troubleshooting with customer technical support, it was found that the basal delivery totals in the total daily dose did not match.It was noted that the basal edit screen was accessed, and the customer made changes to the basal program.The bolus totals did not match as well.The basal history matched the active basal program settings.This complaint is being reported because the patient experienced hypoglycemia associated with use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
kristen lopolito
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key8211132
MDR Text Key131888384
Report Number2531779-2019-00028
Device Sequence Number1
Product Code OYC
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received12/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age30 YR
Patient Weight170
-
-