Model Number 1MTEC30 |
Device Problems
Break (1069); Material Separation (1562)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If implanted, give date: not applicable, as the product is not implantable.If explanted, give date: not applicable, as the product is not implantable.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Event Description
|
It was reported that pieces of a 1mtec30 cartridge came off into the patient's operative eye.The case was finished as normal and the pieces of the cartridge were irrigated out of the operative eye.There was no patient injury and the patient is doing fine post-operatively.Through follow up it was indicated this event may have occurred due to a handling problem.No additional information was provided.
|
|
Manufacturer Narrative
|
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no additional complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Manufacturer Narrative
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
|