|
It was reported that during an unknown procedure, a performer introducer leaked at the diaphragm.There has been no report that any part of the device remained in the patient's body, that the patient required any additional procedures, or that the patient experienced any adverse effects due to this event.Patient demographics, history, and further event details are not available.
|
|
Investigation ¿ evaluation.A review of the dimensional verification, complaint history, device history record, manufactures instructions, quality control, trends, and a functional test & visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned package confirmed that one rycfw-7.0-35-6-rb-cw with the dilator was returned for investigation.The dilator was not inserted into the introducer.There was no visible damage on the introducer or the dilator.The device had visible biomatter present.The tubing was occluded with hemostats and water was injected into the side arm to test for leakage.The device leaked around the silicone disc.The connector cap was disassembled from the check flo body and the cap-to-bottom distance was 1.393¿.There was glue noted in the fitting threads and on top of the fitting, but no glue in the cap.Additionally, cook reviewed the device history record to check for related nonconformances and associated complaints.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should also be noted that the only other complaints associated with either of these lot numbers are all from the same facility.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Measures are being conducted to address this failure mode.Per the quality engineering risk assessment no further action is required.A review of the device history record indicated that the lot met all finished goods release criteria.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|