| Model Number |
86110- |
| Medical Device Problem Code |
False Alarm (1013)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
| Date of Event |
10/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during testing, the procedure could not be made because the valve of the tube with the ball was damaged.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during testing, the procedure could not be made because the valve of the tube with the ball was damaged.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample was received for evaluation.The reported event on cuff won¿t inflate is confirmed.An inflation/deflation test was performed of air was applied to the cuff using a syringe and it was observed inflation was difficult and deflation was hard, it was noticed inside the pilot balloon there is organic material, the pilot balloon was cut and tested by applying air through a syringe and no restrictions were felt, a syringe needle was connected to the inflation line and it was observed the air has difficulties to get through the inflation system in the area where the inflation line is inserted into the tube.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|