| Catalog Number |
01.12.24SN |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
12/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 03 january 2019: lot 184806: 36 items manufactured and released on 02 october 2018.Expiration date: 2023-09-20.No anomalies found related to the issue.To date, 9 items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
During the primary hip surgery and after the stem was implanted, the surgeon noticed a foreign substance on the neck of the stem.The surgeon explanted the size 4 quadra h stem and replaced it with another size 4 quadra h stem.This caused a 5 minute delay in the case and the surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection performed by r&d project manager: visual investigation shows some scratches on the base of the taper cone of the femoral stem and few dot-like scratches on the neck.It is not possible to understand if these are scratches or additional material.Additional information has been requested to the surgeon.Microscopic analysis revealed that there has been a problem during polishing phase and final control.
|
| |
|
Search Alerts/Recalls
|