• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  CANCELLOUS BONE SCREW 6.5X30MM; BONE SCREWS AND PINS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  CANCELLOUS BONE SCREW 6.5X30MM; BONE SCREWS AND PINS : SCREWS Back to Search Results
Catalog Number 117230000
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Pain (1994); Test Result (2695); No Code Available (3191)
Date of Event 08/24/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision due to pain and elevated metal ions.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANCELLOUS BONE SCREW 6.5X30MM
Common Device Name
BONE SCREWS AND PINS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8217280
Report Number1818910-2019-79883
Device Sequence Number1670692
Product Code HWC
UDI-Device Identifier10603295007623
UDI-Public10603295007623
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K970929
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2013
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number117230000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/07/2018
Supplement Date Received by Manufacturer01/16/2019
Initial Report FDA Received Date01/04/2019
Supplement Report FDA Received Date01/31/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-