| Model Number |
ASPC28-3E. |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
When connecting the patient, air was inside the 2 catheter branches.The dialysis session was immediately stopped and the doctors were called.Blood presence was on the dressing.With a visual inspection, it was noticed a cut between the triangle and the catheter part on the patient side.The catheter was removed by the doctor.No clinical consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the investigation performed, the product was manufactured according to required specifications.100% visual and functional (leak test) inspections are performed during the manufacture of the material as detection control; and the fact that product was implanted for 2 years 8 months with no reported problems, it is concluded that a root cause could not be attributable to the manufacturing processes.
|
| |
|