| Catalog Number |
133FV-13 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reports a right side deflation of a tissue expander.The device has been explanted.The replacement device is a second tissue expander.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified white particles.A leak test was perform and two openings were found.A microscopic analysis identified: two curved striated openings on the patch assessed as surgical damage.Based on the device analysis the final assessment is: two curved striated openings assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reports a right side deflation of a tissue expander.The device has been explanted.
|
| |
|