MEDTRONIC HEART VALVES DIVISION MELODY TRANSCATHER PULMONARY VALVE; PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number PB1018
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Endocarditis (1834); Insufficiency, Valvular (1926); Valvular Stenosis (2697); No Information (3190)
Date of Event 11/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that approximately three years and nine months after the implant of this transcatheter pulmonary bioprosthetic valve, the valve was replaced.No additional adverse patient effects were reported. .
 
Additional Manufacturer Narrative
Additional information was received that approximately two years and three months after implant, aggregatibacter arophilus endocarditis was diagnosed and 6 weeks of intravenous antibiotics were prescribed.Approximately two years and five months after implant, balloon dilation of the valve was performed and more than mild regurgitation was noted.Progressive stenosis was noted.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.A review of the sterility lot record confirms that the biological indicators (bi) tested negative for both tissue and solution.This event the occurrence of endocarditis was greater than 12 months post implant (2+ years).Based on the descriptive comments outlined from the journal literature, complaints which occurred greater than 12 months post implant were largely considered to be community acquired (refer to note 1).Therefore, it was unlikely that the endocarditis originally came from the device and/or manufacturing valve process.Endocarditis related risks are addressed in the current risk management file.The reported stenosis and regurgitation may have due to the endocarditis.Note 1 - mylonakis, e., and calderwood, s.B.Infective endocarditis in adults.New england journal of medicine.345 (18).1318-1330.Nov.1, 2001.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MELODY TRANSCATHER PULMONARY VALVE
Common Device Name
PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key8218694
Report Number2025587-2019-00062
Device Sequence Number16558370
Product Code NPV
UDI-Device Identifier00643169415874
UDI-Public00643169415874
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P140017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/07/2016
Device Model NumberPB1018
Device Catalogue NumberPB1018
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2018
Supplement Date Received by Manufacturer01/22/2019
02/25/2019
Initial Report FDA Received Date01/04/2019
Supplement Report FDA Received Date02/18/2019
03/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age14 YR
Patient Weight57
-
-