| Model Number |
U41502H22 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
|
| |
|
Event or Problem Description
|
|
It was reported that upon removal from the packaging, frayed material was allegedly observed on the pta balloon.Another device was used to complete the procedure.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a manufacturing review was performed.The lot met all release criteria.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found frayed material near the proximal cone and distal tip.Therefore, the investigation is confirmed for the reported frayed material.The definitive root cause for the identified frayed material could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review:the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that upon removal from the packaging, frayed material was allegedly observed on the pta balloon.Another device was used to complete the procedure.There was no reported patient contact.
|
| |
|