MEDTRONIC, INC. OPEN PIVOT MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 500DM25
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Information (3190)
Date of Event 12/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that this 25mm mitral mechanical valve was implanted and explanted during the same procedure.The device was replaced with a 24mm mechanical valve for unknown reasons.No additional adverse patient effects were reported. .
 
Additional Manufacturer Narrative
Correction submitted to clarify the event description: medtronic received additional information that this 25mm mitral mechanical valve was implanted and explanted during the same procedure.The reason for the replacement is due to the significant scar tissue on the ventricular side, which caused one of the leaflets to not move freely.As a result the valve was removed and successfully replaced with a 24mm mechanical valve.No additional adverse patient effects were reported.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic received information that this 25mm mitral mechanical valve was implanted and explanted during the same procedure.Due to the patients non medtronic bioprosthetic valve being stenosed, which caused this 25mm valve not to seat properly.As a result the valve was removed and successfully replaced with a 24mm mechanical valve.No additional adverse patient effects were reported.  updated device code.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN PIVOT MECHANICAL HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
MDR Report Key8220646
Report Number3008592544-2019-00001
Device Sequence Number14113638
Product Code LWQ
UDI-Device Identifier00643169335042
UDI-Public00643169335042
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P990046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/27/2021
Device Model Number500DM25
Device Catalogue Number500DM25
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2018
Supplement Date Received by Manufacturer01/24/2019
02/15/2019
Initial Report FDA Received Date01/06/2019
Supplement Report FDA Received Date01/30/2019
02/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient Weight77
-
-