STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
|
Back to Search Results |
|
| Catalog Number |
5530-G-311 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
12/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding revision due to instability involving a triathlon insert was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: no medical records were received for review with a clinical consultant.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event itself could not be confirmed and exact cause of the event could not be determined because insufficient information was provided.Further information such as product return, pre and post operative x-rays, operative reports as well as patient history and follow up notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Event: surgeon determined knee was unstable.He removed femoral component and insert.He implanted ts femoral component with thicker insert.Patient stable.No complications.
|
| |
|