Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that a piece of the grasper remained inside of the patient.Please note that the surgeon elected not to perform the xray.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: grapser broke, missing the pin.Probable root cause: poor autoclave reliability.Shear pin failure in handle.Use error.The reported failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that a piece of the grasper remained inside of the patient.Please note that the surgeon elected not to perform the xray.
|
| |
|