Model Number CQF7574
Medical Device Problem Codes Deflation Problem (1149); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.(expiration date 08/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure of a declot the pta balloon allegedly ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation; the user's locking flow switch was also returned.The device was prepped per the instructions for use, and inflated with a 10ml syringe.The user's locking flow switch was also used during functional testing, and no issues were identified locking the device in the inflated position.The device was deflated within the specified time constraints, without any issues.Therefore, the investigation is unconfirmed for the reported deflation issue.The definitive root cause for the reported deflation issue could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a declot the pta balloon allegedly had difficulty deflating.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8224600
Report Number2020394-2018-02289
Device Sequence Number207203
Product Code LIT
UDI-Device Identifier00801741060496
UDI-Public(01)00801741060496
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF7574
Device Catalogue NumberCQF7574
Device Lot NumberRECV2523
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2019
Initial Date Received by Manufacturer 12/12/2018
Supplement Date Received by Manufacturer03/14/2019
Initial Report FDA Received Date01/07/2019
Supplement Report FDA Received Date03/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient Weight44
-
-