| Catalog Number |
03.632.055 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier reported as (b)(4).The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.A device history record review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a transforaminal lumbar interbody fusion (tlif) on (b)(6) 2018, several instruments failed.When inserting a titanium cannulated screw into lumbar four (l4), the ratchet t-handle made a clicking sound and slipped against the t15 driver shaft.The surgeon determined it would not work and requested a different handle.The shaft was changed to another t-handle and surgeon completed inserting the screws just fine.Then, at the end of the case, the surgeon selected a transconnector and began locking it using the standard torque limiting handle and straight t15 driver shaft.The handle clicking as normal and torquing out, the driver shaft broke off into the recess of the transconnector.The surgeon removed the driver shaft and transconnector then requested another driver shaft and transconnector.Other items with same part number were obtained and tried again.This time, instead of the tip breaking, the tip of the shaft began to deform.Surgeon requested a third shaft and the torque limiter clicked as it should on all three (3) screws of the transconnector.Surgery was completed with a the three to five (3-5) minute delay to gather secondary instruments.There were no complications to the patient.Concomitant devices: screws (part: unknown, lot: unknown, quantity: 3); torque limiting handle (part: 03.632.204, lot: unknown, quantity: 1); titanium cannulated matrix screw (part: 04.616.745, lot: unknown, quantity: 1); t25 stardrive shaft for matrix (part: 03.632.003, lot: unknown, quantity: 1).This report is for a straight tip t15 driver- standard.This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 1 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual inspection: the returned device was examined, and the distal tip was found to be broken.The broken tip was found stuck in ti snap-opn transconnector.The observed damage is consistent with the reported complaint condition as such the complaint condition can be confirmed.No further visual defects or deficiencies were noted.Dhr review: the returned part 03.632.055 was manufactured in october,2018 at haegendorf manufacturing site.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Material analysis: according to dhr, the material was reviewed and the hardness value was confirmed to meet the specification with no non-conformance noted.Conclusion: a definitive root cause could not be determined based on the provided information.This complaint is confirmed however no product design issues or manufacturing discrepancies that would contribute to the reported complaint condition were identified during this investigation.No new, unique or different patient harms were identified because of this evaluation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot part: 03.632.055, lot: l954834, manufacturing site: haegendorf, release to warehouse date: 11.Oct.2018.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.The material was reviewed and the hardness value was confirmed to meet the specification with no non-conformance noted.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|