• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC STRAIGHT TIP T15 DRIVER- STANDARD; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC STRAIGHT TIP T15 DRIVER- STANDARD; SCREWDRIVERS Back to Search Results
Catalog Number 03.632.055
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier reported as (b)(4).The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.A device history record review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a transforaminal lumbar interbody fusion (tlif) on (b)(6) 2018, several instruments failed.When inserting a titanium cannulated screw into lumbar four (l4), the ratchet t-handle made a clicking sound and slipped against the t15 driver shaft.The surgeon determined it would not work and requested a different handle.The shaft was changed to another t-handle and surgeon completed inserting the screws just fine.Then, at the end of the case, the surgeon selected a transconnector and began locking it using the standard torque limiting handle and straight t15 driver shaft.The handle clicking as normal and torquing out, the driver shaft broke off into the recess of the transconnector.The surgeon removed the driver shaft and transconnector then requested another driver shaft and transconnector.Other items with same part number were obtained and tried again.This time, instead of the tip breaking, the tip of the shaft began to deform.Surgeon requested a third shaft and the torque limiter clicked as it should on all three (3) screws of the transconnector.Surgery was completed with a the three to five (3-5) minute delay to gather secondary instruments.There were no complications to the patient.Concomitant devices: screws (part: unknown, lot: unknown, quantity: 3); torque limiting handle (part: 03.632.204, lot: unknown, quantity: 1); titanium cannulated matrix screw (part: 04.616.745, lot: unknown, quantity: 1); t25 stardrive shaft for matrix (part: 03.632.003, lot: unknown, quantity: 1).This report is for a straight tip t15 driver- standard.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is report 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual inspection: the returned device was examined, and the distal tip was found to be broken.The broken tip was found stuck in ti snap-opn transconnector.The observed damage is consistent with the reported complaint condition as such the complaint condition can be confirmed.No further visual defects or deficiencies were noted.Dhr review: the returned part 03.632.055 was manufactured in october,2018 at haegendorf manufacturing site.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Material analysis: according to dhr, the material was reviewed and the hardness value was confirmed to meet the specification with no non-conformance noted.Conclusion: a definitive root cause could not be determined based on the provided information.This complaint is confirmed however no product design issues or manufacturing discrepancies that would contribute to the reported complaint condition were identified during this investigation.No new, unique or different patient harms were identified because of this evaluation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot part: 03.632.055, lot: l954834, manufacturing site: haegendorf, release to warehouse date: 11.Oct.2018.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.The material was reviewed and the hardness value was confirmed to meet the specification with no non-conformance noted.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT TIP T15 DRIVER- STANDARD
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8226702
Report Number2939274-2019-55752
Device Sequence Number1910264
Product Code HXX
UDI-Device Identifier10705034719115
UDI-Public(01)10705034719115
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.632.055
Device Lot NumberL954834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2019
Initial Date Received by Manufacturer 12/12/2018
Supplement Date Received by Manufacturer01/03/2019
01/30/2019
02/06/2019
Initial Report FDA Received Date01/08/2019
Supplement Report FDA Received Date01/30/2019
01/31/2019
02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-