SMITH & NEPHEW, INC. BLADE,3.5 BONE CUTTER FULL RADIUS,DISP; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
7210486 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the procedure, the device had shavings that were peeling off.The metal pieces were suctioned out.A backup device was available to complete the procedure with no delay and no patient injuries.
|
| |
|
Additional Manufacturer Narrative
|
|
One unused 3.5mm full radius bonecutter blades was returned for evaluation.Visually, the part exhibited some signs of tarnish on the inner edgeform.There was no plating issues noted on the inner blade when reviewed under magnification.Functional assessment of the blade confirmed the inner blade rotated freely within the outer blade in the unloaded condition.Silver plating is an smith+nephew process used at the time of manufacture of this blade to improve lubricity and allow for friction free contact between the stainless steel inner and outer blades.The appearance of the "tarnish" and discoloration does not have any adverse effect on the performance of the device, procedure or patient.The tarnishing discoloration is a cosmetic issue only.Without the return of the device that exhibited the reported shedding this investigation could not identify any evidence of product contribution to the reported complaint.Reportedly the bone shavings were suctioned from the surgical site and there was no harm to the patient.Since no harm is being reported and the procedure was completed with a backup.
|
| |
|