| Model Number |
21133-4 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/23/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) of the last three shipments of the plate 21133-4 and the resp.Screws that were send to (b)(6) were inspected and showed no deviations.The quality forms met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing that would contribute to this complaint condition.A review of the raw material device history record revealed no deviation.The material was determined to be conforming and was used as is per product development approval.The proximal plate holes showed only minor deformation.Thus the locking screws were not fully engaged in the plate.Furthermore, the screws backed out due to non-union of the humeral fracture.
|
| |
|
Event or Problem Description
|
|
It was reported that screws backed out of a proximal humeral plate, 4-hole postoperatively.Original implant date (b)(6) 2016.A revision surgery was performed on (b)(6) 2016.
|
| |
|