| Catalog Number |
01.12.031 |
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6) 2019: lot 050833: 24 items manufactured and released on (b)(6) 2005.Expiration date: (b)(6) 2010.No anomalies found related to the problem.To date, all the items of this same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
Revision surgery performed 13 years after primary due to peri-prosthetic metallosis on the proximal part of the stem and on the periphery of the non-medacta cup.No invasion of the joint capsule.Liner and head were ceramic implants.The surgeon revised the stem with a quadra r size 1.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection performed by r&d project manager visual inspection performed on 23rd september 2019.The ha coating of the stem was almost totally absorbed and lots of scratches were visible in the neck region, probably some occurred during the revision; anyway, it is not clear if some of these scratches are related to impingement with other products, as the competitor's cup geometry is unknown.From the received piece it is not possible to determine with certainty the root cause of the event, but it could be related to the competitor's cup.
|
| |
|