| Model Number |
M00573020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an jagtome rx was used in duodenum during a sphincterotomy procedure performed on (b)(6) 2018.According to the complainant, during procedure, first incision was successfully performed using the device.However, during the second attempt of incision, the cutting wire broke.Reportedly, no part was detached inside the patient.The procedure was completed with a second jagtome rx, using the same generator and active cord.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ok.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).Device codes: the problem code 1069 captures the reportable event of cutting wire broken.Visual examination of the returned device revealed that the cutting wire was broken and blackened.There was no other issue noted.The complaint was consistent with the reported event of cutting wire broke.It is most likely that a peak of voltage could have caused the failures noted.Therefore, the most probable root cause is adverse event related to procedure since it is most likely that due to procedural factors encountered during the procedure the performance of the product was limited.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release to distribution.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an jagtome rx was used in duodenum during a sphincterotomy procedure performed on (b)(6) 2018.According to the complainant, during procedure, first incision was successfully performed using the device.However, during the second attempt of incision, the cutting wire broke.Reportedly, no part was detached inside the patient.The procedure was completed with a second jagtome rx, using the same generator and active cord.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ok.
|
| |
|