| Model Number |
9733065 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The stem was returned to the manufacturer.Functional testing found that the component would not fit into the base.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used for a cranial biopsy.It was reported that the site opened 2 navigus biopsy kits and neither on was able to lock the guide stem into the straight base.On the kit they ended up using an angled base and that resolved the issue.There was a reported delay of less than one hour and no other reported impact to patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: the hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|