• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number 09-1462-0056
Medical Device Problem Code Sparking (2595)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/20/2018
Type of Reportable Event Malfunction
Event or Problem Description
Prior to the procedure a spark was seen when the amplifier was powered on.There were no adverse consequences to the user.The procedure was rescheduled.
 
Additional Manufacturer Narrative
One ep-workmate scu was received for evaluation.During inspection the returned device would not turn on and the reported sparking was confirmed.The power supply was temporarily replaced with a test standard power supply and normal power was restored.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.The reported event of a spark was consistent with a non-functional power supply.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported spark could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL
Common Device Name
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key8233188
Report Number2184149-2019-00004
Device Sequence Number1910465
Product Code DQK
UDI-Device Identifier05415067003126
UDI-Public05415067003126
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number09-1462-0056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/20/2018
Supplement Date Received by Manufacturer02/18/2019
03/29/2019
Initial Report FDA Received Date01/10/2019
Supplement Report FDA Received Date02/19/2019
04/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-