| Model Number |
09-1462-0056 |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Prior to the procedure a spark was seen when the amplifier was powered on.There were no adverse consequences to the user.The procedure was rescheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
One ep-workmate scu was received for evaluation.During inspection the returned device would not turn on and the reported sparking was confirmed.The power supply was temporarily replaced with a test standard power supply and normal power was restored.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.The reported event of a spark was consistent with a non-functional power supply.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported spark could not be determined.
|
| |
|