C.R. BARD, INC. (COVINGTON) -1018233 PROXIS¿ URETERAL ACCESS SHEATH Back to Search Results
Model Number 231245
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 12/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that there was a burr on the inner cease.When the user found it, the surgery was completed and no additional device was used.
 
Additional Manufacturer Narrative
The reported event was confirmed but the cause was unknown.The evaluation of the returned sample was performed by futurematrix interventional.The dilator was inserted into the sheath and no resistance was noted.Visual inspection, with the unaided eye, did show a burr or flash inside the opening of the tip of the sheath.Approximately 1 inch of the sheath was cut away for better investigation of the tip.This piece of the sheath was inspected under 25x magnification.The small burr or flash appeared to be peeling away from the inside of the sheath tip.Futurematrix interventional concluded that the sheath tipping process was 100% inspected for defects such as this, and therefore it could not be determined if the event occurred during the manufacture or with use.A possible root cause could be that the flash peeled away during use of instrumentation during the procedure.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "description: the proxis¿ ureteral access sheath is a two component ureteral dilation system which contains a single lumen for injection of fluids as well as passage of endoscopes and related instruments.The packaged product includes the following items: 1 - hydrophilic-coated dilator with female luer connector 1 - hydrophilic-coated sheath with hub indications for use: the proxis¿ ureteral access sheath is indicated for use in endoscopic urology procedures where ureteral dilation and ureteral access is desired for injection of fluids and insertion and removal of endoscopes and related instruments.Contraindications: ¿ patients who are contraindicated for retrograde urological procedures.¿ patients who are contraindicated for antegrade urologic procedures, including but not limited to patients with blood clotting anomalies due to coagulopathies or pharmacological anticogagulations.¿ patients who have the presence of tight strictures which would limit use of the device.¿ patients who have the presence of large obstructing distal ureteral calculi.Warning: for single use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/ or lead to device failure, which in turn, may result in patient injury, illness or death.Reuse reprocessing or re-sterilization may also create a risk of contamination of the device and/or cause patient infection or crossinfection, including, but not limited to, the transmission of infectious disease(s) from one patient to another.Contamination of the device may lead to injury, illness or death of the patient.Do not use if sterile barrier is damaged.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and with applicable laws and regulations.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Precautions: the recommendations given are meant to serve only as a basic guide to the utilization of this access sheath.The ureteral access sheath should not be used without comprehensive knowledge of the indications, techniques and risks of the procedure.To minimize resistance during advancement, ensure the hydrophilic coating on the dilator and sheath is activated with saline or sterile water prior to placement.Do not advance sheath without the dilator in place as this could lead to trauma or injury to patient.Do not bend the device prior to placement as this could damage the integrity of the device and result in patient injury.Do not advance or withdraw device if any resistance is encountered during placement or removal without determining cause and taking action.Adverse events potential adverse events associated with the use of the transurethral access device include, but are not limited to: - mucosal irritation, inflammation and edema - urethral strictures - acute bleeding or hemorrhage - urethral, bladder, or ureteral perforation - other injury to the urinary tract directions for use: 1.Activate the hydrophilic coating by placing the dilator and sheath components into saline or sterile water.Place an 0.035" (0.889mm) or 0.038" (0.965mm) guidewire into the ureter using standard endourology techniques.2.Ensure the dilator lock is securely engaged with sheath hub prior to insertion.3.Insert the guidewire into the tapered end of the dilator/sheath assembly and gradually advance the assembly into the ureter.Note: placement of the assembly can be verified using fluoroscopy or radiographic means.4.While maintaining sheath position, disengage the dilator lock from the sheath hub to gently remove the dilator.Do not advance sheath without the dilator in place.Note: suture holes are provided on sheath hub for securing externally, if desired.5.An endoscope and/or related instruments can now be used through the ureteral sheath as needed.6.If desired, irrigation can be applied using the luer connector on the dilator.7.Upon completion of the access procedure, gently withdraw the device.8.Discard the device in accordance with hospital procedures and with applicable laws and regulations.For urological use only".
 
Event or Problem Description
It was reported that there was a burr on the inner cease.When the user found it, the surgery was completed and no additional device was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIS¿ URETERAL ACCESS SHEATH
Common Device Name
PROXIS
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8233339
Report Number1018233-2019-00209
Device Sequence Number1745797
Product Code FED
UDI-Device Identifier00801741101724
UDI-Public(01)00801741101724
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number231245
Device Catalogue Number231245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2019
Initial Date Received by Manufacturer 12/17/2018
Supplement Date Received by Manufacturer04/18/2019
Initial Report FDA Received Date01/10/2019
Supplement Report FDA Received Date05/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-