Catalog Number FS-LXB-3X6
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional concomitant medical products: boston scientific alliance handle, unknown model, medwork cable and handle, unknown model.Occupation: non-healthcare professional.Investigation evaluation: the product is being returned for evaluation and the investigation is on-going.A follow-up emdr will be provided within 30 days of submission of this report with product evaluation information.
 
Event or Problem Description
During a lithotripsy procedure, the physician used a cook fusion lithotripsy extraction basket.An extraction [of stones was attempted] with the basket.A distant learning [sic] was not possible without lithotripsy.The lithotripsy extraction basket handle [was loaded] into the alliance handle.A slow attempt to shred the stone [was made].The lithotripsy extraction basket was torn by the hand [handle].Emergency lithotripsy was not possible.Surgery was performed to remove the stones and the lithotripsy extraction basket.A section of the device did not remain inside the patient¿s body.The patient required surgery to remove the stones and basket due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Boston scientific alliance handle, unknown model.Medwork cable and handle, unknown model.Cook fusion omni-tome pre-loaded sphincterotome, fs-omni-35-260.Occupation: non-healthcare professional.Investigation evaluation: our evaluation of the product said to be involved confirmed the report.The orange wire guide lumen on the distal end of the device has been stripped along its full length.The catheter has kinks throughout the length.The wire basket appeared stretched and was returned separated from the drive wire, likely cut.The drive wire was returned separate from the catheter and not attached to the handle.It appears as if the drive wire broke near the handle.The drive wire was 183 cm in length.There was brown/red residue on the inside tip of the basket and along some parts of the drive wire.The handle advanced and retracted without resistance.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use (ifu) state the following warning: "basket wires may fragment and/or stone impaction may occur during lithotripsy, requiring surgical intervention." the ifu states the following: "caution: if difficulty is encountered when removing basket from duct, do not use excessive force." the ifu also states "damage to wire guide lumen may result if zip port is visible and device is pulled from accessory channel with wire guide locked in wire guide locking device." prior to distribution, all fusion lithotripsy extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a lithotripsy procedure, the physician used a cook fusion lithotripsy extraction basket.An extraction [of stones was attempted] with the basket.A distant learning [sic] was not possible without lithotripsy.The lithotripsy extraction basket handle [was loaded] into the alliance handle.A slow attempt to shred the stone [was made].The lithotripsy extraction basket was torn by the hand [handle].Emergency lithotripsy was not possible.Surgery was performed to remove the stones and the lithotripsy extraction basket.A section of the device did not remain inside the patient¿s body.The patient required surgery to remove the stones and basket due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION LITHOTRIPSY EXTRACTION BASKET
Common Device Name
FFL, DISLODGER, STONE, BASKET, URETHRAL, METAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key8236278
Report Number1037905-2019-00015
Device Sequence Number1624678
Product Code FFL
UDI-Device Identifier00827002482784
UDI-Public(01)00827002482784(17)211022(10)W4134748
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/22/2021
Device Catalogue NumberFS-LXB-3X6
Device Lot NumberW4134748
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Initial Date Received by Manufacturer 12/19/2018
Supplement Date Received by Manufacturer12/19/2018
Initial Report FDA Received Date01/10/2019
Supplement Report FDA Received Date02/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS TJF-180 ENDOSCOPE.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-