• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION CG FUTURE BAND; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION CG FUTURE BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 638RL38
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Information (3190)
Date of Event 12/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that immediately post implant of this 38mm mitral annuloplasty ring, the ring was explanted and replaced with a non-medtronic bioprosthetic valve.The reason for replacement was not reported.No additional adverse patient effects were reported. .
 
Additional Manufacturer Narrative
Medtronic received additional information that after repairing the valve, the surgeon elected to replace the valve instead.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key8236368
Report Number2025587-2019-00140
Device Sequence Number13729903
Product Code KRH
UDI-Device Identifier00643169793576
UDI-Public00643169793576
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K061127
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/17/2023
Device Model Number638RL38
Device Catalogue Number638RL38
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/03/2019
Supplement Date Received by Manufacturer01/09/2019
01/04/2019
Initial Report FDA Received Date01/10/2019
Supplement Report FDA Received Date01/28/2019
03/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient Weight60
-
-