| Model Number |
620RG27 |
| Medical Device Problem Codes |
Gradient Increase (1270); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Calcium Deposits/Calcification (1758); Mitral Regurgitation (1964); Mitral Valve Stenosis (1965); No Information (3190)
|
| Date of Event |
01/15/2013
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 2 years 10 months post implant of this 27mm mitral annuloplasty ring, it was explanted and replaced with a 27mm mitral bioprosthetic valve for unknown reasons.No additional adverse patient effects were reported. .
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that approximately two years and eight months post implant of this annuloplasty ring, echocardiogram revealed a mean gradient of 7 mmhg and an exercise gradient of 16 mmhg.Approximately two months following, cardiac catheterization identified an increased mean gradient of 13 mmhg which peaked over 26 mmhg with exercise.Ultimately two years and ten months post implant of the annuloplasty ring, the ring was explanted and replaced due to severe stenosis.During the explant procedure, significant calcification in the trigone areas of the anterior leaflet and along the base of the posterior annulus was identified.The physician indicated much of the valve orifice obstruction was due to the subvalvular apparatus.No additional adverse patient effects were reported. .
|
| |
|