U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION RING 620RG27 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION RING 620RG27 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY Back to Search Results
Model Number 620RG27
Medical Device Problem Codes Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Calcium Deposits/Calcification (1758); Mitral Regurgitation (1964); Mitral Valve Stenosis (1965); No Information (3190)
Date of Event 01/15/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 2 years 10 months post implant of this 27mm mitral annuloplasty ring, it was explanted and replaced with a 27mm mitral bioprosthetic valve for unknown reasons.No additional adverse patient effects were reported. .
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that approximately two years and eight months post implant of this annuloplasty ring, echocardiogram revealed a mean gradient of 7 mmhg and an exercise gradient of 16 mmhg.Approximately two months following, cardiac catheterization identified an increased mean gradient of 13 mmhg which peaked over 26 mmhg with exercise.Ultimately two years and ten months post implant of the annuloplasty ring, the ring was explanted and replaced due to severe stenosis.During the explant procedure, significant calcification in the trigone areas of the anterior leaflet and along the base of the posterior annulus was identified.The physician indicated much of the valve orifice obstruction was due to the subvalvular apparatus.No additional adverse patient effects were reported. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RING 620RG27 DURAN ANCORE ANNULOPLASTY C
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key8236634
Report Number2025587-2019-00143
Device Sequence Number1148247
Product Code KRH
UDI-Device Identifier00721902859786
UDI-Public00721902859786
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K032810
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2013
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/27/2012
Device Model Number620RG27
Device Catalogue Number620RG27
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/07/2019
Supplement Date Received by Manufacturer04/22/2019
Initial Report FDA Received Date01/10/2019
Supplement Report FDA Received Date04/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-