Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a female patient underwent a breast augmentation procedure with a mentor tissue expander 700cc device that deflated intraoperatively.The expander for the left breast failed to connect to the valve, causing leakage and making it impossible to use.As a result, the device had to be explanted from the patient.No information regarding a replacement device has been received.
|