• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 366444
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that six bd vacutainer® sst¿ ii advance plus blood collection tubes experienced under filling during use.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.However, bd has initiated further investigation relating to this issue through a capa and potential causes have been identified.As a result, corrective actions have been established and are in the process of being implemented.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.However, further investigation activities have been conducted through a capa and the most likely root cause has been identified.As a result, corrective actions and procedures are being implemented to mitigate further occurrences.Root cause description: although no samples or photos were available for evaluation, bd has initiated further investigation through a capa.The capa has identified the most likely root causes and corrective actions are in the process of being implemented.Rationale: based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue.The investigation has identified potential root cause(s) for this issue and corrective actions are in the process of being implemented.
 
Event or Problem Description
It was reported that six bd vacutainer® sst¿ ii advance plus blood collection tubes experienced under filling during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
MDR Report Key8241012
Report Number9617032-2019-00024
Device Sequence Number16760125
Product Code JKA
UDI-Device Identifier50382903664448
UDI-Public50382903664448
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number366444
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 08/30/2017
Supplement Date Received by Manufacturer08/30/2017
Initial Report FDA Received Date01/11/2019
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-