|
(b)(4).Batch # r92r3j.Device analysis: the analysis results found that the b11lt instrument was received with the sleeve melted.This type of damage is consistent with the one produced by an energized device.The device was visually inspected and no damaged was found on the clear lens.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|