According to the reporter, during procedure, the dissector alarmed red during use.When the dissector was removed from patient, the surgical technologist wiped off the blade which resulted in the blade breaking off from dissector.Nothing fell into the patient's cavity.The surgeon opened a new device to complete without incident.There was no patient injury.
|
|
Evaluation summary: one device was returned for evaluation.The returned product did not meet specification as received.Visual inspection of the disposable device revealed that the tip of the waveguide was broken off.The broken piece was returned with the device.The reported condition was confirmed.Investigation personnel performed functional testing on the returned hand piece using a test lab generator and battery.The assembled device returned a green light and a welcome tone, but immediately returned a red light emitting diode (led) indicator with an error tone (alarm activation) when the device was activated.This characteristic indicated that the device was not functional.The waveguide had its tip missing, but the missing piece was returned with the device.Investigation personnel concluded that the titanium waveguide cracked from continued use after becoming damaged and may have come in contact with any of the following; hemostats, clips, staples, retractors, etc.During use.The investigation identified the cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.Contact with metal objects during use will cause the active blade to crack and may eventually break off.This issue is specific to ultrasonic dissectors.The instructions for use (ifu) advises device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use dam aged components.Use of damaged components may result in injury to the patient or user.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|