| Model Number |
ETBF3616C145E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Death (1802); Low Blood Pressure/ Hypotension (1914); Rupture (2208)
|
| Date of Event |
12/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other relevant device(s) are: product id: etlw1616c82e, serial/lot #: (b)(4), ubd: 06-apr-2016, udi#: (b)(4); product id: etlw1616c124e, serial/lot #: (b)(4), ubd: 05-nov-2016, udi#: (b)(4); product id: etcf3636c49e, serial/lot #: (b)(4), ubd: 06-apr-2016, udi#: (b)(4); product id: etlw1616c93e, serial/lot #: (b)(4), ubd: 10-nov-2018, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 65mm in diameter abdominal aortic aneurysm.At implant, the proximal aortic neck diameter was 20-23-31-36mm and 15mm in length.The neck angulation was 74 degrees.Heli-fx endoanchors were prophylactically implanted during the index procedure.It was reported approximately 4 years post the index procedure the patient presented with abdominal pain and hypotension.Aneurysm rupture was confirmed.Surgical intervention was completed however the patient expired due to the ruptured aortic aneurysm one day post the intervention.Per the investigator the cause of the event was related to the endograft device.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products updated to include the following devices: etlw1610c93es; s/n: (b)(4); use by date: 23-nov-2019; upn # (b)(4).Etlw1616c93e ;s/n: (b)(4); use by date: 02-jan-2018; upn # (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to outcomes attributed to adverse event: death and intervention.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|