• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM CANNULATED SCREW PARTIALLY THREADED/54MM; SCREW,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM CANNULATED SCREW PARTIALLY THREADED/54MM; SCREW,FIXATION,BONE Back to Search Results
Model Number 214.554
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, and the product evaluation is in progress.No conclusion can be drawn.Additionally, device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the surgeon was repairing a medial epicondyle fracture of the distal humerus.During the procedure, the thread of the first cannulated screw unraveled.While the surgeon was inserting second cannulated screw by hand over a guide ware, the tip of the screw broke and remained in the patient.Surgeon recovered both screws and the threads of one of the screws which basically raveled, uncoiled out of the shaft.It is unknown if there was surgical delay.Procedure outcome and patient status is unknown.Concomitant devices: guide wire (part 292.72, lot unknown, quantity 1) , washer (part 219.910, lot l804246, quantity 1) , screws (part unknown, lot unknown, quantity 3).This report is for one (1) cannulated screw.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluation: broken visual inspection: upon visual inspection, it was observed that the 4.5 mm partially threaded cannulated screw/54 mm was broken at its tip.The broken tip fragment was not received with the device and was retained in the body of the patient.No surface wear was observed on the screw.The received condition of the device was determined to agree with the complaint condition dimensional inspection: due to post-manufacturing damage, dimensional inspection could not be performed.During the document/specification review the following drawings, reflecting the current revision, were reviewed.-- cannulated screw 4.5 xx*/xx hex sd no device history record (dhr) review, including material review, could be conducted as the lot number of the device is unknown.As the received condition was determined to agree with the complaint condition, this complaint is confirmed.Investigation conclusion: the complaint condition was confirmed as the 4.5 mm partially threaded cannulated screw/54 mm was determined to be broken through visual inspection.No new product design issues or manufacturing discrepancies were identified during this investigation.Based upon the investigation findings, no additional corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM CANNULATED SCREW PARTIALLY THREADED/54MM
Common Device Name
SCREW,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8244751
Report Number2939274-2019-55814
Device Sequence Number13941441
Product Code HWC
UDI-Device Identifier10886982154231
UDI-Public(01)10886982154231
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K963172
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number214.554
Device Catalogue Number214.554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Initial Date Received by Manufacturer 12/18/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date01/14/2019
Supplement Report FDA Received Date02/06/2019
Patient Sequence Number1
Patient Age14 YR
-
-